Hioki 100:1 PROBE, 10:1 PROBE 9666, 9665

Voltage testing options for Memory HiCorders

The 9666 and 9665 are 100:1 and 10:1 attenuation probes used to measure high voltage with Hioki Memory HiCorders.

Note: This probe does not expand the maximum rated voltage above ground of an isolated input.
*Please check for Memory HiCorder compatibility by referring to the recorder's product catalog.

Attenuates the input voltage to 1/100 or 1/10
• Input capacitance to the probe can be lowered to reduce the load effect

Basic specifications

  9666 9665
Division ratio 100 : 1 ±1% 10 : 1 ±1%
Maximum input voltage 5 kV peak (1 MHz or less) 1 kV rms (500 kHz or less)
Max. rated voltage to earth 1000 V CAT II * 1000 V CAT II *
  * Maximum rated voltage above ground depends on input unit. This probe does not expand the maximum rated voltage above ground of an isolated input.
Input Impedance 100 MΩ 10 MΩ
Frequency characteristics DC to 200 MHz DC to 150 MHz
Input capacitance 7 pF or less 18 pF ±2 pF
Compensation capacitance range (Equipment side) 15 to 45 pF 15 to 45 pF
Cable length 1.5 m (4.92 ft)
Accessories Vinyl case ×1